Additel 835
Portable Calibration Bath
Price on request
The ADT835 is a portable calibration bath with a compact bath size of 3.54 x 3.54 inches (90 x 90 mm), setting a new benchmark for powerful, space-saving calibration solutions.
- Product description
- Extra information
- Video
- Downloads
-
The ADT835 is a portable calibration bath with a compact bath size of 3.54 x 3.54 inches (90 x 90 mm), setting a new benchmark for powerful, space-saving calibration solutions. Unlike typical integrated dry-liquid baths, the ADT835 delivers exceptional versatility—accommodating temperature sensors of virtually any type, size, or shape. Its innovative design supports both batch and field calibrations with ease, while maintaining excellent stability, uniformity, and accuracy. Combining the precision and reliability of a traditional calibration bath with the portability of a dry well, the ADT835 is a dependable choice for demanding calibration environments.
The ADT835 delivers exceptional performance through a smart, user-focused design. Its advanced data logging capabilities simplify documentation and help minimize errors or missed readings. Built-in application tools enhance workflow efficiency, while intelligent remote control functionality allows for multitasking and increased productivity. The 6.5-inch TFT touchscreen provides an intuitive interface, making operation straightforward—even for first-time users.
• Engineered for exceptional portability and performance
• Extended temperature range: -30°C to 250°C
• High-precision accuracy: ±0.1°C
• Compatible with a wide range of probe types and sizes
• Optional built-in process calibrator
• Quick to temperature
• Delivers outstanding stability and uniformity
• Fast thermal response for efficient operation
• Features advanced dual-zone heating technology
• One touch self-calibration
• Designed with robust electromagnetic interference protection -
General Specifications
Specification Description
Power requirementsPower supply
110V: 100~120V, 50 Hz/60Hz, fuse: T8A 250V
220V: 200~230V, 50 Hz/60Hz, fuse: T4A 250V
Maximum power consumption: 800W
Size and weightMain structure size: 7.5 × 16.2 × 15.2 in (191(W)×412(H)×387(D)mm)
Maximum size: 7.5 × 18.1 × 16.77 in (191(W)×460(H)×426(D)mm) (with handle, overflow tube)
Weight: 35.27 Ibs (16 kg) /ADT835-155 and 35.94 Ibs (16.3 kg ) /ADT835PC-155
30.86 Ibs (14 kg )/ADT835-250 and 31.53 Ibs (14.3kg) ADT835PC-250
EnvironmentOperating temperature: 0°C~ 40°C Guaranteed accuracy: 13°C~ 33°C Storage temperature: -20°C~ 60°C Humidity: 5%RH~95%RH, non-condensing Altitude: < 2000 m
IP rating: IP20, indoor use only
Vibration: 2g Impact: 5g
Package drop test: 1m
Warm-up time: 15 minWorking area Bath size: 90mm*90mm
Working area: bath centerφ75mm circular zone, 15mm above bath bottom, 65mm below the fluid level, maximum depth
155mmHeat transfer fluid Maximum volume: 2.5L
Recommended fluid at 150°C: DOW CORNING XIAMETER PMX-200 10cSt Recommended fluid at 250°C: DOW CORNING XIAMETER PMX-200 50cStCommunication USB-A, USB-B, LAN, WIFI, BLE Electrical functions Reference RTD channel (EXT.REF)
φ4mm banana jacks, 6 PIN Lemo port, quick push wiring DUT channels (CH1, CH2)
φ4mm banana jacks, quick push wiringDisplay 6.5 in (165 mm) color capacitive touch screen, 640*480, anti-reflective panel, Dark or bright theme switchable
Communication update rate: 10 times/s, Display refresh rate: 4 times/s Temperature units: °C, °F, K
Highest resolution: 0.001°C, adjustableLanguages English, Chinese, Japanese, Russian, German, French, Italian, and Spanish Warranty 1 year Temperature specification[1] Models ADT835PC-150 / ADT835-150ADT835PC-250 / ADT835-250Range[2] -30°C~150°C33°C~250°CINT REF accuracy[3] ± 0.1°C± 0.1°CEXT REF accuracy [4] ± 0.05°C± 0.05°CStability[5] ± 0.01°C± 0.015°CUniformity[6] ± 0.02°C± 0.03°C (≤200°C)
± 0.04°C ( > 200°C)Repeatability[7] 0.04°C (internal reference)
0.01°C (external reference)0.04°C (internal reference)
0.01°C (external reference)Heating time[8] 18 min (-30~23°C)30 min (23~150°C)20 min (23~100°C)
35 min (100~250°C)Cooling time[8] 45 min (150°C~23°C)
45 min (23°C~-25°C)
55 min (23°C~-30°C)30 min (250°C~100°C)
45 min (100°C~33°C)Typical time to stability [9] 10 min10 min[1] The following temperature specifications are all based on the following conditions:When the environmental temperature is between 13°C~33°C and the calibration bath is within one year after calibration, the confidence level is 95% (K=2), the bath is not open, the working area is theφ75mm circular zone of the bath center, and it is 15mm above the bath bottom and 65mm below the fluid level. The depth of the fluid should be more than 130mm.When the environmental temperature exceeds the range of 13°C~33°C, the stability and uniformity specification will be 1.25 times larger.When the bath is open, the stability and uniformity specification will be 1.25 times larger.[2] The lower limit of controlling range can vary depending on the environmental temperature.[3] The internal reference accuracy has taken into account the followingl sources of error including calibration uncertainty, stability, uniformity and repeatability.[4] The external reference accuracy is available for ADT835PC-150 and ADT835PC-250, the reference probe used is Accumac model AM1730-12-ADT.[5] Stabilty is defined as half of the difference between the maximum and minimum temperature variations of the heat transfer fluid within the working zone over a 10-minute period after the product has reached full stabilization.[6] Uniformity is defined as half of the difference between the highest temperature point and the lowest temperature point in the valid working area.[7] Repeatability mainly considers the hysteresis and short-term stability of the reference sensor.[8] Heating and cooling times are measured from the start of temperature control to when the setpoint is reached within the internal accuracy error band. Actual times may vary depending on ambient temperature, AC power supply voltage, sensor load, and whether the bath is covered during operation.[9] The typical time to stability refers to the time taken for the temperature in the bath to reach the stability tolerance after the heating or cooling time ends.
Input Specifications (Process Calibrator [PC] Option)
Electrical specification Reference RTD channel (EXT. REF) RTD measureRange: (0~400)ΩAccuracy: ±1.25mΩ@(0~50)Ω, ±0.0025%RD@(50~400)ΩResolution: 0.1mΩTemperature coefficient: ±1ppmF.S/°C@(0~13)°C and (33~50)°CExcitation current: 0.75mA, current reversalMeasurement technology: 4-wire, 1S measure cycleType: Pt25, Pt100 SPRTStandard type: ITS-90, CVD, IEC-751Wiring methods: Lemo and quick push connectorDUT channels
(CH1, CH2)Current measureRange: (-30~30) mAAccuracy: ±(0.01%RD+2μA)Resolution: 0.1μATemperature coefficient: ±5ppm F.S/°C@ (0~13)°C and (33~50)°CInput impedance: <10ΩVoltage measureRange: (-30~30)V, (-12~12)VAccuracy: ±(0.01%RD+0.6mV)Resolution: 0.1mVTemperature coefficient: ±5ppmF.S/°C@ (0~13)°C and (33~50)°CInput impedance: >1MΩSwitch measureMechanical switch or electrical switchLoop power 24 V±10%(with Load), 24 V±1%(no Load), Max 60mA for supply power to temperature transmitter and electronic thermometer HARTCH1 supports measure and adjust HART transmitterRTD measureRange: (0~400)Ω and (0~4000)Ω(auto)Accuracy: ±2.0mΩ@(0~25)Ω, ±0.004%RD@(25~400)Ω, ±0.005%RD@(400~4000)Ω (4-wire measure)Resolution: 0.1mΩTemperature coefficient: ±2ppmF.S/°C@( 0~13)°C and (33~50)°CExcitation current: 0.25mA, current reversalMeasurement technology: 2/3/4-wire, 1S measure cycleRTD type: Pt10, Pt25, Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Cu100, Ni100, Ni120Wiring methods: Lemo and quick push connectorTC measureSignal range: (-75~75)mVAccuracy: ±(0.01%RD+5μV)Temperature coefficient: ±5ppmF.S/°C @(0~13)°C and (33~50)°CTC type: S, R, B, K, N, E, J, T, C, D, G, L, UCold junction range: 0°C~ 50°CCold junction accuracy: ±0.2°CNote: This specification reflects the bath’s accuracy for up to one year following calibration, based on a confidence level of K=2
Temperature sensor measurement specification Channel Type Measurement Accuracy Reference RTD channel (EXT.REF) PT25 Temp. (°C) -30050100150200250Accuracy (°C) ±0.013±0.013±0.013±0.013±0.013±0.014±0.014PT100 Temp. (°C) -30050100150200250Accuracy (°C) ±0.006±0.006±0.008±0.009±0.011±0.012±0.013DUT channels
(CH1, CH2)PT100 Temp.(°C) -30050100150200250Accuracy (°C) ±0.009±0.010±0.012±0.015±0.017±0.019±0.021K-TC Temp. (°C) -30050100150200250Accuracy (°C) ±0.14±0.13±0.13±0.13±0.14±0.15±0.15N-TC Temp. (°C) -30050100150200250Accuracy (°C) ±0.20±0.19±0.19±0.18±0.17±0.17±0.17E-TC Temp. (°C) -30050100150200250Accuracy (°C) ±0.09±0.09±0.08±0.08±0.08±0.09±0.09J-TC Temp. (°C) -30050100150200250Accuracy (°C) ±0.11±0.10±0.10±0.10±0.11±0.11±0.11T-TC Temp. (°C) -30050100150200250Accuracy (°C) ±0.14±0.13±0.12±0.12±0.11±0.11±0.11Note 1: The measurement accuracy specification does not include the accuracy of the DUT itself;Note 2: Thermocouple measurement accuracy is calculated based on the cold junction temperature of 0°C and does not include the error of the CJC sensor.
Models ADT835PC-150ADT835-150ADT835PC-250ADT835-250Range -30°C~150°C-30°C~150°C33°C~250°C33°C~250°CmA/mV/V ••measure ••Switch measure ••24V loop power ••HART communicator ••Quick push wiring ••INT reference ••••EXT reference ••Self-calibration ••Documenting task ••Step test ••••Thermal calculator ••••Sensor library ••••Snapshot ••••Remote control ••••Smart diagnosis ••••USB stick update ••••
-
Documents:
Filename Upload date Size Additel 835 Datasheet 2025.6.25 1.7 MB