DPI 620 Genii IS
Multifunction calibrator, with Atex approval
Price on request
The Druck DPI 620 Genii IS is an easy to use, rugged and high accuracy multifunction test and calibration system with optional full HART and Foundation Fieldbus communication for hazardous areas. It is the IS version of the highly successful DPI 620 GENII and replaces the DPI620 IS calibrator. It brings all the benefits of the DPI620 GENII with ATEX and IECEx approval for zones 1 and 2.
Anderen kochten ook:
- Product description
- Extra information
- Video
- Downloads
-
The Druck DPI 620 Genii IS is an easy to use, rugged and high accuracy multifunction test and calibration system with optional full HART and Foundation Fieldbus communication for hazardous areas. It is the IS version of the highly successful DPI 620 GENII and replaces the DPI620 IS calibrator. It brings all the benefits of the DPI620 GENII with ATEX and IECEx approval for zones 1 and 2.
It’s GE’s next-generation instrument for the calibration, configuration and maintenance of electrical, frequency, pressure and temperature devices in measurement and control applications. DPI 620 Genii IS can measure and source a wide range of parameters simultaneously including mA, mV, V, Ohms, frequency, RTDs, T/Cs, and HART/Fieldbus digital signals. It includes a loop power supply to energize devices and control loops and a stabilized DC voltage supply for ratio metric transducers. It has improved easy to detach battery pack that allows hot swopping in hazardous areas and includes an LED charge indicator to show battery condition without the need to turn the device on. With its versatility and power, it’s the only tool you’ll need for instrument and loop maintenance across demanding multi process parameters.
Main features
- Multi-function: electrical, frequency, temperature and pressure parameters
- 5 Channels for simultaneous measurement and sourcing of mA, V, Ohms, frequency, pressure, RTDs and T/Cs, HART, Fieldbus, `
- ATEX and IECEx approval for hazardous zones 1 and 2
- Simple and user friendly touch screen interface.
- Quick launch menu applications such as DASHBOARD, TASK selection & FAVORITES
- 1 year uncertainty for mA 0.016%RDG + 0.0065%FS
- 1 year uncertainty for pressure from 0.025%FS
- 31 plug ‘n play pressure modules from 25 mbar to 1000 bar (PM620 option)
- Quick pressure generation from vacuum to 100 bar pneumatic or 1000 bar hydraulic (PV621/PV622/PV623 options)
- Full HART, Foundation Fieldbus & Profibus communicator option
- Fully documenting – automated procedures, PASS/FAIL errors, results storage
- Interface to GE calibration software or other leading calibration software packages
- Data logging up to 32Gb storage. Enough for a complete database of procedures & results
- Hand held unit compact and lightweight - only 1.1 Kg/2.4lb.
- Intrinsically safe Lithium-Polymer battery provides reliable power for a full 8hr shift
-
Intrinsically safe DPI620G-IS general specifications for hazardous area use
Processor and memory
800 MHz Processor
512 MB 800 MHz SDRAM
8 GB Internal flash memory
Display
Size: 110 mm (4.3 in) diagonal; 480 x 800 pixels
LCD: Colour display with touch-screen
Protected by 2 mm toughened glass, impact tested in accordance with BS EN 61010-1:2010 (0.5kg object from 1 m)
File Viewers
A Windows® desktop is available for managing files, running third party applications and viewing simple images, Word documents, Excel spreadsheets, PDF files and PowerPoint files
Languages
English {Default}, Chinese, French, German, Italian, Portuguese, Russian, Spanish, Dutch, Japanese
Operating temperature
-10° to 50°C (14° to 122°F)
Storage temperature
-20° to 70°C (-4° to 158°F)
Ingress Protection
IP55
Humidity
0 to 90% RH Non condensing
Shock / Vibration
BS EN 61010-1:2010; MIL-PRF-28800F for Class II equipment, 1 m Drop Tested
EMC
Electromagnetic compatibility: BS EN 61326-1:2006
Electrical safety
Electrical – BS EN 61010-1: 2010
Pressure safety
Pressure Equipment Directive - Class: Sound Engineering Practice (SEP)
Approval
CE Marked
Baseefa 16ATEX0002X
Ex II 1 G
Ex ib IIC T4 Gb (-10°C ≤ Ta ≤ +50°C)
IECEx BAS 16.0010X
Size (L: W: H)
183 x 114 x 55 mm (7.2 x 4.5 x 2.2 in)
Weight
1.1 kg (2.4 lb) – battery included.
Power supply
Lithium-ion battery (GE Part number: DPI620G-IS-BATTERY); Capacity: 4800 mAh. Nominal voltage: 3.75 V. Charge temperature: 0° to 40°C (32° to 104°F) Discharge temperature: -10° to 50°C (14° to 12°F). Charge/discharge cycles: > 500 > 70% capacity. Safe area charging only using external charger DPI620G-IS-CHARGER and universal mains adaptor IO620-PSU. The battery is detached from the instrument using two thumb screws and mounted on the charger. The battery can be taken into a hazardous area without being connected to an instrument and can attached and detached in the be hazardous area. The battery has an LED indicator to show the charge state of the battery without having to turn the instrument on or when it is not attached to an instrument
Duration
Measure functions (CH1): ≈ 7 hours continuous. Dual Function, mA measure (CH2): ≈ 5 hours (24 V Source at 12 mA)
Connectivity
USB Type Mini B (client)
Electrical Measurement and Source NLH&R1
±1°C (2°F) for 24 hrs
(note 2)Total Uncertainty
10° to 30°C (50° to 86°F)
for 1 year (note 3)Additional error
-10° to 10°C (14° to 50°F)
30° to 50°C (86° to 122°F)Resolution Display
reading
window%Rdg
+%FS
%Rdg +%FS %Rdg/
°C+%FS/
°CMeasure mode
DC Voltage
Thermocouple
Please refer to Thermocouple specification table
CH1 TC mode -10 to 100 mV
0.0045
0.008
0.007 (0.009)
0.01
0
0.0005
0.001
CH1
+/- 200 mV
0.0045
0.004
0.01
0.005
0
0.0005
0.001
CH1
CH2
+/- 2000 mV
0.004
0.003
0.0095 (0.01)
0.005
0
0.0005
0.01
CH1
CH2
+/- 20 V
0.0025
0.002
0.0145
0.002
0
0.0005
0.00001
CH1
CH2
+/- 30 V
0.0035
0.0035
0.0145
0.004
0
0.0005
0.0001
CH1
CH2
AC Voltage
(note 1)
Not applicable to DPI620G-IS
versions0 to 2000 mVAC
0.125
0.125
0.2
0.15
0.005
0.005
0.1
CH1
0 to 20 VAC
0.1255
0.125
0.2
0.15
0.005
0.005
0.001
CH1
0 to 300 VAC
1
0.06
1.5
0.1
0.05
0.005
0.01
CH1
Current
+/- 20 mA
0.006
0.005
0.012 (0.016)
0.006 (0.0065)
0
0.0005
0.0001
CH1
CH2
+/- 55 mA
0.005
0.005
0.016 (0.019)
0.005 (0.006)
0
0.0005
0.0001
CH1
CH2
Resistance (True, 4 wire)
RTD
Please refer to RTD specification table
CH1 0 to 400 Ω
0.0055 (0.006)
0.001 (0.002)
0.009
0.0012
0
0.0005
0.001
CH1
0 to 4000 Ω
0.0055 (0.006)
0.001 (0.002)
0.009
0.0012
0
0.0005
0.01
CH1
Resistance (4 wire)
RTD
Please refer to RTD specification table
CH1 0 to 400 Ω
0.012
0.005
0.015
0.006
0
0.001
0.001
CH1
0 to 4000 Ω
0.0115
0.0045
0.015
0.006
0
0.001
0.01
CH1
Frequency
0 to 1000 Hz
0.0003
0.0002
0.003
0.0002
0.0001
CH1
1 kHz to 50 kHz (5 kHz)
0.0003
0.0004
0.003
0.0004
0.00001
CH1
0 to 999999 CPM
Refer to range table above for equivalent frequency
0.01
CH1 0 to 999999 CPH
Refer to range table above for equivalent frequency
0.01
CH1 Trigger level
Automatic and adjustable 0 to 20 V
0.1
Trigger level
Automatic or manual setting 0 to 20 V
0.1
Pressure
25 mbar to 1000 bar
Please refer to PM 620 pressure range table
P1 P2 IDOS external module
Please refer to IDOS UPM datasheet. Cable P/N IO620-IDOS-USB required
IDOS USB port
Please refer to GE Sensing for compatible devices
USB Source mode DC Voltage
TC mode
Please refer to Thermocouple specification table
TC mode -10 to 100 mV
0.009
0.008
0.014
0.01
0
0.0005
0.001
CH1
0 to 200 mV
0.0045
0.004
0.01
0.005
0
0.0005
0.1
CH1
0 to 2000 mV
0.004
0.003
0.009
0.005
0
0.0005
0.1
CH1
0 to 20 V (12 V) @ 3 mA
max.
0.006
0.002 (0.0035)
0.0145 (0.015)
0.002 (0.004)
0
0.0005
0.001
CH1
Current
0 to 24 mA with ext. loop power
0.01
0.004
0.015
0.005
0
0.0005
0.001
CH1
CH2
0 to 24 mA with int. loop power
0.01
0.004
0.015
0.005
0
0.0005
0.001
CH2
Internal loop power
15V +/-10%; 100R output impedance
Resistance2
RTD
Please refer to RTD specification table
CH1 0 to 400 Ω (0.1 mA)
0.024(0.026)
0.0035 (0.0045)
0.03 (0.035)
0.0075 (0.012)
0
0.001
0.01
CH1
0 to 400 Ω (0.5 mA)
0.004
0.0025
0.008
0.003
0
0.001
0.01
CH1
400 to 2000 Ω (0.05 mA)
0.048
0.0035
0.06
0.006
0
0.001
0.01
CH1
2 k to 4 kΩ (0.05 mA)
0.048
0.0035
0.06
0.0045
0
0.001
0.01
CH1
Maximum input current
0-400 Ω 5 mA, 400-2000 Ω 1mA, 2000-4000 Ω 0.5 mA
Frequency
0 to 1000 Hz
0.0003
0.00023
0.003
0.00023
0.1
CH1
1 kHz to 50 kHz (5 kHz)
0.0003
0.000074
0.003
0.000074
0.001
CH1
Output waveform
Square, positive swing up to 20 V (12V) adjustable, negative swing -120 mV (fixed)
Sine and Triangular, adjustable amplitude and offset within the limits -2.5 (-0.5) to +20 V (+12 V)Square wave peak output
0 to 20 V +/-20 mV (3 mA maximum)
0 to 99999 CPM
Please refer to range table above for equivalent frequency
1 CH1 0 to 99999 CPH
Please refer to range table above for equivalent frequency
1 CH1 Notes:
Values in () apply to DPI620G-IS models
1. Specification applies, 45 to 65 Hz and between 10% and 100% of full scale
2. Specification applies when calibration temperature is between 10 and 30°C
3. Total uncertainty includes reference standard uncertainty, NLH&R and typical long term stability for one year (K=2)Multiple parameter display capability
The display can be configured to show a maximum of 6 (5 for IS versions) simultaneous reading windows as follows: CH1, CH2, P1, P2, IDOS (not IS versions), HART/Fieldbus)
“True Ohms” RTD Measure Mode (4-wire)
Type
Temperature coefficient
Temperature range
Total Uncertainty
10° to 30°C (50° to 86°F)
for 1 year°C
°F
Rdg
Tos
From
To
From
To
%
°C
°F
Pt 50
3.85
-200
850
-328
1562
0.012
0.05
0.09
Pt 100
3.85
-200
850
-328
1562
0.012
0.04
0.07
Pt 100
3.92
-200
850
-328
1562
0.012
0.04
0.07
Pt 200
3.85
-200
260
-328
500
0.01
0.03
0.051
260
850
500
1562
0.15
0.077
0.14
Pt 500
3.85
-200
-60
-328
-76
0.01
0.026
0.044
-60
0
-76
32
0.015
0.05
0.086
0
850
32
1562
0.012
0.05
0.086
Pt 1000
3.85
-200
-150
-328
-238
0.009
0.024
0.04
-150
0
-238
32
0.011
0.036
0.061
0
850
32
1562
0.012
0.036
0.061
Cu 10
4.27
-200
0
-328
32
0
0.14
0.25
0
260
32
500
0
0.17
0.3
D 100
6.18
-200
0
-328
32
0.01
0.035
0.06
0
640
32
1184
0.012
0.035
0.06
Ni 100
6.72
-60
0
-76
32
0
0.026
0.047
0
250
32
482
0
0.03
0.055
Ni 120
6.72
-80
0
-112
32
0
0.022
0.04
0
270
32
518
0
0.028
0.05
270
320
518
608
0
0.057
0.1
Standard RTD Measure Mode (4-wire)
Type
Temperature coefficient
Temperature range
Total Uncertainty
10° to 30°C (50° to 86°F)
for 1 year
°C
°F
Rdg
Tos
From
To
From
To
%
°C
°F
Pt 50
3.85
-200
850
-328
1562
0.021
0.16
0.28
Pt 100
3.85
-200
0
-328
32
0.017
0.1
0.175
0
850
32
1562
0.0215
0.1
0.174
Pt 100
3.92
-200
0
-328
32
0.017
0.1
0.175
0
850
32
1562
0.0215
0.1
0.174
Pt 200
3.85
-200
0
-328
32
0.017
0.069
0.12
0
260
32
500
0.018
0.069
0.12
260
850
500
1562
0.033
0.33
0.6
Pt 500
3.85
-200
-60
-328
-76
0.0165
0.051
0.09
-60
0
-76
32
0.017
0.16
0.29
0
850
32
1562
0.024
0.16
0.28
Pt 1000
3.85
-200
-150
-328
-238
0.016
0.044
0.074
-150
0
-238
32
0.018
0.1
0.175
0
850
32
1562
0.0215
0.1
0.174
Cu 10
4.27
-200
0
-328
32
0.035
0.66
1.18
0
260
32
500
0.01
0.66
1.18
D 100
6.18
-200
0
-328
32
0.019
0.1
0.174
0
640
32
1184
0.02
0.1
0.174
Ni 100
6.72
-60
0
-76
32
0
0.071
0.13
0
250
32
482
0.002
0.071
0.13
Ni 120
6.72
-80
270
-112
518
0
0.06
0.11
270
320
518
608
0
0.2
0.36
RTD Simulate Mode (0.1 mA min, 0-400 Ω; 0.05 mA min, 400-4000 Ω)
Type
Temperature coefficient
Temperature range
Total Uncertainty
°C
°F
Rdg
Tos
From
To
From
To
%
°C
°F
Pt 50
3.85
-200
850
-328
1562
0.043 (0.052)
0.24 (0.35)
0.42 (0.63)
Pt 100
3.85
-200
850
-328
1562
0.04 (0.047)
0.16 (0.22)
0.28 (0.40)
Pt 100
3.92
-200
850
-328
1562
0.04 (0.047)
0.16 (0.22)
0.28 (0.40)
Pt 200
3.85
-200
260
-328
500
0.0345 (0.041)
0.12 (0.16)
0.21 (0.29)
260
850
500
1562
0.087
0.28
0.50
Pt 500
3.85
-200
-60
-328
-76
0.33 (0.038)
0.095 (0.12)
0.169 (0.22)
-60
850
-76
1562
0.078
0.23
0.41
Pt 1000
3.85
-200
-150
-328
-238
0.32 (0.037)
0.085 (0.11)
0.15 (0.20)
-150
260
-238
500
0.0675
0.19
0.34
260
850
500
1562
0.082
0.17
0.31
Cu 10
4.27
-200
260
-328
500
0
0.85 (1.40)
1.53 (2.52)
D 100
6.18
-200
640
-328
1184
0.38 (0.046)
0.16 (0.22)
0.28 (0.40)
Ni 100
6.72
-60
250
-76
482
0
0.12 (0.16)
0.22 (0.29)
Ni 120
6.72
-80
270
-112
518
0
0.11 (0.14)
0.20 (0.25)
270
320
518
608
0
0.25
0.45
Notes:
These figures relate to DPI 620 Genii uncertainties only Values in () apply to DPI620G-IS models
For RTD Measure and Source functions the uncertainty is given by:-
Urtd = T(°C) x %Rdg + Tos (°C)or
Urtd = T(°F) x%Rdg + Tos (°F)
where T( ) is the measurement expressed in °C or °FMeasurement resolution:
0.01 °C/F
Simulation resolution 0.1 °C/FExcitation current:
Measure mode 0 to 400Ω 2.5mA, 400Ω to 4000Ω 0.5 mA;
Simulate mode 0 to 400Ω 5mAmax, 0.4 to 2kΩ 1mA max and 2 to 4 kΩ 0.5mA max
Simulate mode pulsed excitation current minimum duration 10 ms
Thermocouple Measurement and Simulation Measurement Simulation Type Standard Temperature range Total Uncertainty 10° to 30°C (50° to 86°F) for 1 year
°C °F °C °F °C °F From
To
From
To
B
IEC 584
250.00
500.00
482.00
932.00
4.00
7.20
4.00
7.20
500.00
700.00
932.00
1,292.00
2.00
3.60
2.00
3.60
700.00
1,200.00
1,292.00
2,192.00
1.50
2.70
1.50
2.70
1,200.00
1,820.00
2,192.00
3,308.00
1.00 (1.10)
1.80 (1.98)
1.10
1.98
E
IEC 584
-270.00
-200.00
-454.00
-328.00
2.00
3.60
2.00
3.60
-200.00
-120.00
-328.00
-184.00
0.50
0.90
0.50
0.90
-120.00
1,000.00
-184.00
1,832.00
0.25
0.45
0.30
0.54
J
IEC 584
-210.00
-140.00
-346.00
-220.00
0.50
0.90
0.50
0.90
-140.00
1,200.00
-220.00
2,192.00
0.30
0.54
0.40
0.72
K
IEC 584
-270.00
-220.00
-454.00
-364.00
4.00
7.20
4.00
7.20
-220.00
-160.00
-364.00
-256.00
1.00
1.80
1.00
1.80
-160.00
-60.00
-256.00
-76.00
0.50
0.90
0.50
0.90
-60.00
800.00
-76.00
1,472.00
0.30 (0.40)
0.54 (0.72)
0.40
0.72
800.00
1,370.00
1,472.00
2,498.00
0.50
0.90
0.60
1.08
L
DIN 43710
-200.00
-100.00
-328.00
-148.00
0.40
0.72
0.40
0.72
-100.00
900.00
-148.00
1,652.00
0.25
0.45
0.30
0.54
N
IEC 584
-270.00
-200.00
-454.00
-328.00
7.00
12.60
7.00
12.60
-200.00
-40.00
-328.00
-40.00
1.00
1.80
1.00
1.80
-40.00
1,300.00
-40.00
2,372.00
0.40
0.72
0.50
0.90
R
IEC 584
-50.00
360.00
-58.00
680.00
3.00
5.40
3.00
5.40
360.00
1,760.00
680.00
3,200.00
1.00
1.80
1.10
1.98
S
IEC 584
-50.00
70.00
-58.00
158.00
3.00
5.40
3.00
5.40
70.00
320.00
158.00
608.00
1.50
2.70
1.50
2.70
320.00
660.00
608.00
1,220.00
1.10
1.98
1.20
2.16
660.00
1,740.00
1,220.00
3,164.00
1.00 (1.10)
1.80 (1.98)
1.20
2.16
T
IEC 584
-270.00
-230.00
-454.00
-382.00
3.00
5.40
3.00
5.40
-230.00
-50.00
-382.00
-58.00
1.00
1.80
1.00
1.80
-50.00
400.00
-58.00
752.00
0.30
0.54
0.30
0.54
U
DIN 43710
-200.00
-50.00
-328.00
-58.00
0.60
1.08
0.60
1.08
-50.00
600.00
-58.00
1,112.00
0.30
0.54
0.30
0.54
C
0.00
1,600.00
32.00
2,912.00
0.80 (0.90)
1.44 (1.62)
1.00
1.80
1,600.00
2,000.00
2,912.00
3,632.00
1.00 (1.10)
1.80 (1.98)
1.20
2.16
2,000.00
2,300.00
3,632.00
4,172.00
1.40 (1.50)
2.52 (2.70)
1.70
3.06
D
0.00
100.00
32.00
212.00
1.10
1.98
1.10
1.98
100.00
270.00
212.00
518.00
0.80
1.44
0.80
1.44
270.00
1,200.00
518.00
2,192.00
0.60 (0.70)
1.08 (1.26)
0.70
1.26
1,200.00
1,800.00
2,192.00
3,272.00
0.80 (0.90)
1.44 (1.62)
1.00
1.80
Values in () apply to DPI620G-IS models
Measurement resolution 0.01 °C/F
Simulation resolution 0.1 °C/F
Cold Junction (CJ) Uncertainty 0.2°C (0.4°F) in ambientrange 10 to 30°C (50 to 86°F)
Add 0.01° CJ Uncertainty/° outside of this ambient range -
Documents:
Filename Upload date Size DPI 620 Genii IS Multifunction Calibrator Datasheet 2024.3.04 5.3 MB DPI 620 Genii IS Multifunction Calibrator Manual 2019.3.11 4.4 MB DPI 620 Genii IS Firmware Update Information 2018.3.27 0.3 MB